Chapter 6, General Operation: Sequestration
109
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Figure 55, Blood bag adaptor harness
Inspecting the
Installation
Always inspect the disposable set after completing the installation.
1. Inspect all parts of the disposable set and verify that there are no twists,
kinks, or flat spots.
2. Verify that all connections are secure and all appropriate clamps are
closed.
3. Close the device cover, ensuring that no tubing is inadvertently clamped
off.
1.
5.
4.
2.
3.
1. Effluent line connection
2. White, blue and yellow
ratchet clamps
3. Green ratchet clamps
4. Collection bags
5. Air bag