Chapter 6, General Operation: Sequestration
117
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Salvage procedure, or end the procedure.
Figure 62, The Sequestration Cycle Complete message
To end the Sequestration procedure:
1. Touch
End Procedure
. A “Confirm Procedure Complete” message
appears.
2. Touch
End Procedure
to confirm.
The
Records
screen displays the procedure record. If the device detects there
is still fluid in the bowl, it empties the bowl to the blue line. The device then
pumps a small amount of air through the blue line to flush any remaining blood
in the line into the RBC bag. During this Empty phase, a “Purging Blue Line”
message appears in the message area. When the blue line has been fully
purged, a “Procedure Complete” message appears.
Note: Haemonetics does not recommend resuming a Sequestration procedure
after marking it completed.
Note: To begin a new procedure, you must first power down the device and
power it back on.
Note: If you power down the device before fully purging the blue line, power the
device back on, choose to resume the procedure, and touch
End Procedure
.
This flushes the remaining blood in the blue line into the RBC bag.
Note: If the device is powered off without purging the blue line and is powered
on again within six hours with a processing set installed, the device prompts
you to resume the previous procedure or to save the previous procedure and