Chapter 11, Troubleshooting
205
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
305
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•OPEN the white air line clamp.
•CLOSE the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Touch
Continue
to empty the bowl.
306
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Touch
Continue
to empty the bowl.
307
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•OPEN the white air line clamp.
•CLOSE the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Agitate the RBC bag.
3. Touch
Continue
to concentrate.
Table 16, Event Messages
ID #
Event Message Text