Chapter 11, Troubleshooting
195
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
253
Load Disposable
Explanation:
A disposable must be loaded to begin a procedure.
Corrective Action:
Load a disposable and touch
Start Procedure
when ready to begin a procedure.
254
PRP Separation Lost
Explanation:
A device event has caused the centrifuge to stop and the PRP separation to be lost.
The RBCs in the centrifuge should be emptied to the RBC bag and the cycle started over.
Corrective Action:
1. Touch
Empty Bowl
to empty RBCs to the RBC bag.
2. Touch
Conc
to repeat Sequestration with RBCs from the RBC bag.
OR
1. Touch
Empty Bowl
to empty RBCs back to the RBC bag.
2. Touch
Fill
to repeat Sequestration with RBCs from the blood bag.
255
Recentrifuge Delay
Explanation:
The procedure is establishing a platelet layer. The pump will restart after a 30 second delay
when separation is established.
Corrective Action:
Touch
to override the delay.
256
Data Transfer Interrupted
Explanation:
The device was transmitting data to a storage device and the transmission was interrupted.
This could occur if the storage device was removed from the USB port before completing data
transfer or if the connection between the storage device and the USB port was loose.
Export of the data will have to be repeated.
Corrective Action:
1. Ensure the storage device is securely installed in the USB port.
2. Touch
Export
to begin the data transfer.
Table 16, Event Messages
ID #
Event Message Text