Chapter 7, Protocol Settings
131
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
PRP Collection Pump Rate:
The approximate rate at which the pump
turns while collecting fluid during the PRP collection phase
PRP Collection Centrifuge Speed:
The approximate rate at which the
centrifuge turns during the PRP collection phase
PRP Extend Volume:
The volume in milliliters to pump after the effluent
line sensor detects RBCs during the PRP collection phase. Once this
volume is pumped, the PRP collection is complete.
PRP Milk:
Determines if the process of starting and stopping the PRP
collection phase is desired as determined by the programmed
parameters “Milk Vol” and “Milk Delay.”
PRP Milk Volume:
The volume to pump during the PRP collection until
the pump stops and the Milk delay time starts. This setting is only
adjustable if PRP Milk is set to “Yes.”
PRP Milk Delay:
The time the pump will be stopped during the PRP
collection phase before starting to pump the programmed Milk Vol
parameter. This setting is only adjustable if PRP Milk is set to “Yes.”
PPP Return Plasma:
Determines which clamps to open and close at the
beginning of the Empty phase and to account for volume returned. If you
select “Yes,” the PPP is returned to the bowl and then to the RBC bag
along with the RBCs.