Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
141
Recentrifuge Delay
Explanation:
A recentrifuge delay occurred because the centrifuge and pump were stopped when there was
fluid in the bowl.
The centrifuge is coming up to speed and will spin for 50 seconds to re-establish separation in
the bowl. The pump will start after separation is established.
142
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
143
Device Cover Open
Explanation:
The device cover is open. The device cover must be closed and locked for the procedure to
continue.
Corrective Action:
Close the device cover.
If the problem persists, service is required.
144
Pump Communication Failure
Explanation:
An issue has been detected with the pump.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text