Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
126
Bowl Size Mismatch
Explanation:
A Latham bowl was detected, but a 70 mL bowl was selected.
Corrective Action:
If a 70 mL bowl is installed:
1. Check the bowl and chuck adaptor for proper installation in the centrifuge.
2. Clean the optics.
3. Touch
Keep Bowl Size
.
OR
If a Latham bowl is installed:
1. Touch
Change Bowl Size
.
2. Select the new bowl size or scan the processing set barcode.
127
Bowl Size Mismatch
Explanation:
A 70 mL bowl was detected, but a Latham bowl was selected.
Corrective Action:
If a Latham bowl is installed:
1. Check the bowl and chuck adaptor for proper installation in the centrifuge.
2. Clean the optics.
3. Touch
Keep Bowl Size
.
OR
If a 70 mL bowl is installed:
1. Touch
Change Bowl Size
.
2. Select the new bowl size or scan the processing set barcode.