Chapter 11, Troubleshooting
197
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
260
Processing Set Entry Not Recorded
Explanation:
A processing set has already been entered in the Records screen. The device does not allow
the entry of two processing sets unless the centrifuge arm has been opened and closed while
replacing the first processing set. If the first set was entered incorrectly the data for that set can
be edited by going to the Records screen.
Corrective Action:
1. Go to the Records screen to edit the information for the first processing set.
OR
2. Open and close the centrifuge arm to replace the first processing set. A prompt will be
displayed to enter the new processing set information.
261
Partial Bowl
Explanation:
Blood processed using a partial bowl may have a lower hematocrit than blood processed using
a normal full bowl. Because the hematocrit of the bowl contents is lower, there is more
supernatant in the bowl. In order to dilute the larger volume of supernatant, a partial bowl may
require two times the normal saline solution.
Corrective Action:
Select the desired response pads.
To change the wash volume, touch
Cycle Wash Volume
and adjust manually.
262
Recentrifuge Delay
Explanation:
A recentrifuge delay occurred because the centrifuge and pump were stopped when there was
fluid in the bowl.
The centrifuge is coming up to speed and will spin for 50 seconds to re-establish separation in
the bowl. The pump will start after separation is established. After 25 seconds, an override is
provided if it becomes necessary to restart the centrifuge before completing the full delay.
CAUTION: Overriding may cause red cells to spill into the waste bag due to incomplete
separation of the cells in the bowl.
Corrective Action:
To override the delay and restart the pump now, touch
.