208
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
401
Fat Reduction Wash Cycle
Explanation:
The fat reduction and wash phase is in process. During this phase, salvaged blood is moved
into and out of the bowl in order to isolate the fat and remove it from the packed red blood cells.
The normal washing step occurs in between two distinct fat reduction steps, and a secondary
wash will occur at the end of the cycle.
Corrective Action:
In order to end the specialized fat reduction cycle and execute a normal wash cycle, navigate to
the Protocol Settings screen and activate Haemonetics Default settings, or a custom settings
group with Fat Reduction set to off.
If the process has already completed the wash step using the full wash volume, touching
Empty
will empty the bowl immediately, skipping fat reduction for this cycle.
402
Returning Cells to Bowl
Explanation:
The user has elected to stop a fat reduction cycle during Wash. Packed red blood cells are
being returned to the bowl from the reservoir tubing in order to maintain high hematocrit
product. After this brief phase a normal wash cycle will begin and no further fat reduction steps
will occur unless the setting is changed again.
406
Software Update Not Completed
Explanation:
A software update was started but did not complete.
Corrective Action:
Retry the software update.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text