106
Chapter 6, General Operation: Sequestration
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Figure 52, Closing and locking the header arm
Note: A click will be heard when the locking mechanism is completely secured.
Warning: Verify that the outlet port and effluent tubing are free of any
restrictions prior to initiating a procedure. If the outlet port is inadvertently
clamped off, the
bowl rotary seal may become compromised. See
for more information.
Installing the Effluent Line Sensor Tubing
To install the effluent line tubing:
1. Floss the effluent tubing into the effluent line sensor groove.
2. Ensure that the tubing is completely installed in the effluent line sensor.
Hanging the Waste Bag
To hang the waste bag:
1. Hang the waste bag on the pins on the left side of the device.
2. Verify that the waste bag is securely connected to the effluent line.
3. Ensure that the waste bag drain port is completely closed.
1.
2.
3.
4.
1. Header arm
2. Tubing in effluent line
sensor
3. Header arm latch
4. Bowl in centrifuge chuck