Chapter 11, Troubleshooting
183
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
178
Red Line Valve Position Error
Explanation:
An issue has been detected with the red line valve position. The centrifuge must stop before the
cover will unlock and corrective action can be taken.
Corrective Action:
1. Clamp the red, yellow, and blue lines.
2. Open the valve module cover.
3. Check the tubing manifold placement.
4. Check the tubing placement in the valve module.
5. Close and latch the valve module cover.
6. Unclamp the lines.
7. Close the device cover.
8. Touch
Continue
to continue.
If the problem persists, service is required.
179
Blue Line Valve Position Error
Explanation:
An issue has been detected with the blue line valve position. The centrifuge must stop before
the cover will unlock and corrective action can be taken.
Corrective Action:
1. Clamp the red, yellow, and blue lines.
2. Open the valve module cover.
3. Check the tubing manifold placement.
4. Check the tubing placement in the valve module.
5. Close and latch the valve module cover.
6. Unclamp the lines.
7. Close the device cover.
8. Touch
Continue
to continue.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text