Excessive Pressure in Blue Line
Explanation:
The manifold pressure sensor detected high pressure in the blue line during the Empty phase.
The centrifuge must stop before the cover will unlock and corrective action can be taken.
Corrective Action:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks and occlusions.
3. Touch
Continue
to continue.
If the problem persists, service is required.
168
Excessive Pressure in Red line
Explanation:
The manifold pressure sensor detected high pressure in the red line during the Return phase.
The centrifuge must stop before the cover will unlock and corrective action can be taken.
Corrective Action:
1. Ensure the red line clamp is open.
2. Check the tubing for kinks and occlusions.
3. Touch
Continue
to continue.
If the problem persists, service is required.
169
Pneumatics Failure
Explanation:
An issue has been detected with the device's pneumatics system. It is unable to maintain
adequate pressure.
Corrective Action:
Please service the device before continuing use.