82
Chapter 5, General Operation: Cell Salvage
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
1. Open the cover, centrifuge header arm, valve module cover, and pump
platen.
2. Lift the remaining disposable set components from the tub and drape
them over the device with the waste bag on the left side of the device and
the bowl placed loosely in the centrifuge well.
3. Thread the pump tubing around the pump.
4. Install the tubing manifold on the left side of the valve module, pressing it
lightly into place.
5. Floss the tubing into the air detector.
6. Insert the clear tubing and the color-coded lines into the grooves in the
valve module.
7. Close the pump platen.
8. Close and latch the valve module cover.
Figure 37, Tubing harness
Installing the Bowl
To install the bowl:
1.
70 mL bowl only:
Insert the chuck adaptor into the centrifuge well.
Note: The chuck adaptor is NOT disposable and should be saved for
subsequent procedures.
2. Ensure that the lower port of the bowl faces the effluent line sensor.
3. Install the bowl in the centrifuge by carefully pressing down on the
shoulders of the bowl until it is seated securely in the chuck.
2.
3.
1.
4.
1. Tubing in air detector
2. Pump platen lever
3. Pump platen
4. Tubing in valve module
channels