Device Fan Failure
Explanation:
One of the 3 fans in the device has failed during POST, Failure of this fan does not prevent
operation of the device.
Corrective Action:
1. The procedure can be continued, but it is recommended that the device be serviced prior to
its next use.
2. Touch
OK
to continue.
258
Software CRC Error
Explanation:
An issue has been detected with the update files on the USB flash drive and the software
update was not completed successfully.
Corrective Action:
Service is required to resolve this issue.
259
Software Update Interrupted
Explanation:
The software update process was interrupted. This may have been caused by the USB flash
drive becoming partially removed from the USB port while the device was
transferring data.
Corrective Action:
1. Touch
OK
to clear the even message.
2. Ensure the USB flash drive is properly inserted into the USB port.
3. Touch
Update
to retry the update.
4. Do not remove the USB flash drive until the update is complete.
If the problem persists, service is required.