Chapter 11, Troubleshooting
199
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
269
Confirm Procedure Complete
Explanation:
The user had chosen to end the procedure. Ending the procedure will purge the blue line.
Corrective Action:
Touch
End Procedure
to confirm ending the current procedure.
OR
Touch
Resume Procedure
to resume the procedure in progress,
271
Prepare to Resume Procedure
Explanation:
A procedure that was in progress is being resumed. To resume effectively, the bowl, centrifuge
arm, pump platen, manifold valve cover, device cover, and tubing clamps must be in the correct
positions.
Corrective Action:
1. Ensure the bowl is properly placed in the centrifuge.
2. Ensure the centrifuge arm is locked.
3. Ensure the pump platen is closed.
4. Ensure the manifold valve cover is closed.
5. Ensure the device cover is closed.
6. Ensure the tubing clamps are open as required.
7. Touch
OK
to resume the previous procedure.
272
Purging Blue Line
Explanation:
The blue line is being purged to the reinfusion bag. Purging will complete when air is detected
from the bowl.
Corrective Action:
Wait for purging to complete.
Table 16, Event Messages
ID #
Event Message Text