168
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
121
Centrifuge Arm was Unlatched
Explanation:
The device detected that the centrifuge header arm latch was opened. Was the bowl size or
processing set changed?
Corrective Action:
If the bowl size or processing set was not changed, touch
No
.
If the bowl size or processing set was changed:
1. Touch
Yes
. The device will display the Bowl Selection screen.
2. Follow the prompts on the screen.
122
Fluid Not Detected When Expected
Explanation:
The bowl optics have not detected fluid in the bowl when expected.
Corrective Action:
1. Ensure the red line clamp is open.
2. Check the tubing for kinks and occlusions.
3. If using external vacuum, ensure the level is not greater than 250 mmHg.
4. Check for proper bowl size.
5. If the setup is correct, touch
Continue
to continue.
OR
6. If the bowl size is incorrect, touch
Change Bowl Size
and then select correct bowl size from
the Bowl Selection screen.
NOTE:
When using regulated external suction, ensure that the A&A line is clamped while
setting the regulator vacuum level to the desired level. If the A&A line is not clamped the
vacuum level may exceed the selected vacuum level and affect performance of the device.
Table 16, Event Messages
ID #
Event Message Text