20
Chapter 1, Introduction
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Note: Store disposables in a dry place away from solvent vapors and extremes
of temperature.
Electrical
Specifications
The electrical specifications for operating the Cell Saver Elite device are as
follows
Caution:
The Cell Saver Elite device must be operated in an environment
compatible to the requirements of the IEC/EN 60601-1-2:2001 Standard,
Electromagnetic compatibility (EMC).
Additional IEC/EN compliance information is available in
.
Note: The power source used must be properly grounded.
*In accordance with IEC/EN 60601-1 standard, medical electrical equipment,
general requirements for safety.
Table 4, Environmental Specifications
Conditions
Values
Ambient operating
temperature
10 °C to 27 °C (50 °F to 80.6 °F)
Storage/transportation
temperature
-20 °C to 50 °C (-4 °F to 122 °F)
Operating humidity level
8 to 80% R.H., non-condensing above 0
°C
Atmospheric pressure range
< 2438 meters (8000 ft.)
Table 5, Electrical Input Power
Rated Voltage
Rated Current
Fuse
Frequency
100–120 V
3.0 A
T3.15A250V
50/60 Hz
200–240 V
1.5 A
T3.15A250V
50/60 Hz
Table 6, Enclosure/Chassis Leakage Current Specifications*
Condition
Polarity
Ground
Max Value
Normal
Normal
Normal
100 µ
Α
Reverse
Normal
100 µ
Α
Single fault
Reverse
Open
500 µ
Α
Normal
Open
500 µ
Α