Pneumatics Failure
Explanation:
An issue has been detected with the device's pneumatics system. It is unable to maintain
adequate pressure. Processing cannot continue.
Corrective Action:
Please service the device before continuing use.
227
Air Detected During Fill
Explanation:
The air detector sensed air in the red line during the Fill phase. The blood bag may be empty.
Corrective Action:
If the blood bag is not empty:
1. Ensure the red line clamp is open.
2. Check the tubing for kinks, occlusions, and correct placement in the air detector and pump.
3. Touch
Continue Using Blood Bag
to proceed.
If the blood bag is empty:
1. Replace the blood bag.
2. Touch
Continue Using Blood Bag
to proceed.
OR
To complete the Sequestration cycle:
1. Touch
End Cycle
to end the current cycle.
2. The device will empty the bowl contents to the RBC bag.
Table 16, Event Messages
ID #
Event Message Text