Chapter 1, Introduction
21
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Suction
Specifications
The specifications for the Cell Saver Elite suction are as follows.
Laser
Specifications
The Cell Saver Elite device is a class 3R laser product.
The laser specifications for the Cell Saver Elite device are as follows:
Table 7, Suction Specifications
Characteristics
Values
SmartSuction
Recommended reservoir volume
≤
3 L
Recommended A&A line length
≤
12 ft [3.6 m]
Recommended A&A line inner
diameter
0.3 in [7.6 mm]
Recommended suction tip inner
diameter
0.3 in [7.6 mm]
Operating vacuum
20 to 150 mmHg
(2.7 to 20.0 kPa; 26.7 to 200 mbar)
Vacuum cutoff
175 mmHg
(23.3 kPa; 233 mbar)
Maximum free air flow
40 L/min
Manual Suction
Operating vacuum
50 to 250 mmHg
(6.7 to 33.3 kPa; 66.7 to 333.3
mbar)
Maximum free air flow
40 L/min
Post-Op Suction
Operating vacuum
25 to100 mmHg
(3.3 to 13.3 kPa; 33.3 to 133.3
mbar)
Maximum free air flow
40 L/min
Table 8, Laser Specifications
Characteristics
Values
Max radiation output
3 mW
Wavelength
650 nm
Max light output
7 mW (bowl optics)
1.7 mW +/- 0.2 mW (barcode reader)
Standards
IEC/EN 60825-1:2007
a