Chapter 11, Troubleshooting
189
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
228
Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. Sequestration can continue, but the
device cannot monitor the effluent from the bowl during PRP collection.
Corrective Action:
To proceed with the user monitoring the effluent:
1. Touch
Continue
to continue.
2. Monitor the effluent from the bowl.
3. When ready to end the PRP collection, touch
Empty
to empty the bowl.
4. Follow the device prompts to continue.
OR
To complete the Sequestration cycle:
1. Touch
End Cycle
to end the current cycle.
2. The device will empty the bowl contents to the RBC bag.
If the problem persists, service is required to fully resolve this issue.
229
Post-Op Suction Leak Detected
Explanation:
The device has detected a leak in the suction pathway.
Corrective Action:
1. Check for proper placement of the wound drain.
2. Ensure all connections are secure, including the wound drain, post-op line, reservoir, and
vacuum line.
3. Ensure the spare reservoir ports are securely capped.
4. Check the reservoir for leaks.
If the problem persists, external suction should be used and service is required to resolve this
issue.