Chapter 11, Troubleshooting
167
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
117
Long Empty
Explanation:
The air detector did not sense air when expected, indicating that the device may have pumped
more than the expected volume of fluid from the bowl or that a tubing occlusion could be
preventing fluid from emptying as expected.
Corrective Action:
1. Ensure that fluid is not transferring from the waste bag to the bowl, which would indicate loss
of sterile air.
NOTE:
If fluid is transferring from the waste bag to the bowl, waste may have reached the RBC
bag. The contents of the RBC bag should be returned to the bowl to be washed again.
2. Check the tubing for correct placement in the air detector.
3. Check the effluent tubing for correct placement in the effluent line sensor.
4. Check the effluent tubing for kinks and occlusions.
5. Check the blue and red tubing for kinks and occlusions.
NOTE:
If a kink or occlusion is found in the blue tubing, it is recommended to QC the RBC
product to ensure no hemolysis occurred.
6. Touch
Continue
to continue.
118
Barcode Reader Failure
Explanation:
An issue has been detected with the barcode reader. Scanning has been disabled.
Corrective Action:
1. Enter the information manually using the Records screen.
2. Touch
OK
to close this message.
Service is required to repair the barcode reader.
119
Barcode Reader Failure
Explanation:
An issue has been detected with the barcode reader. Scanning has been disabled.
Corrective Action:
1. Enter information manually using the Records screen.
2. Touch
OK
to close this message.
Service is required to repair the barcode reader.
Table 16, Event Messages
ID #
Event Message Text