130
Chapter 7, Protocol Settings
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Off:
The only option is to end the procedure immediately.
Smart Empty
(On/Off)
Determines if the device slows the pump speed at a decremental rate to empty
fluid from the bowl or if the pump speed stays at a constant speed, adjustable
by the user.
On:
The pump empties fluid from the bowl at a preprogrammed
decremental speed determined by the parameter settings.
Off:
The pump empties fluid at a constant speed adjustable by the user.
Parameters
Cell Salvage
Fill Start Volume:
The approximate volume of fluid in the reservoir that
transitions the device from Standby to Fill when the bowl is empty
Fill Resume Volume:
The approximate volume of fluid in the reservoir
that transitions the device from Standby to Fill when the bowl already
contains some fluid
Fill Pump Rate:
The approximate rate at which the pump turns while
filling the bowl with fluid
Minimum Wash Volume:
The minimum volume of wash solution to be
used to wash one bowl of fluid
Wash Pump Rate:
The approximate rate at which wash solution enters
the bowl
Empty Pump Rate:
The approximate rate at which the pump turns while
emptying fluid from the bowl. This setting is only adjustable if Smart
Empty is set to “Off.”
Emergency Fill Pump Rate:
The approximate rate at which the pump
turns while filling the bowl with fluid in Emergency mode
Emergency Wash Pump Rate:
The approximate rate at which wash
solution enters the bowl while in Emergency mode
Emergency Empty Pump Rate:
The approximate rate at which the
pump turns while emptying fluid from the bowl in Emergency mode. This
setting is only adjustable if Smart Empty is set to “Off.”
Sequestration
Fill Pump Rate:
The approximate rate at which the pump turns while
filling the bowl with fluid and collecting PPP
Fill Centrifuge Speed:
The approximate rate at which the centrifuge
turns while the bowl is filling with whole blood and during the PPP
collection phase