104
Chapter 6, General Operation: Sequestration
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Hanging the RBC Bag
To install the RBC bag:
1. Remove the RBC bag and tubing from the tub and hang the bag on the
top hooks of the right IV pole.
2. Close the two small ratchet clamps on the reinfusion lines.
3. Ensure that the two large ratchet clamps on the blue line are open and the
twist-lock connection is secure.
Figure 49, RBC bag
Installing the Tubing Harness
To install the processing set tubing harness:
1. Open the cover, centrifuge header arm, valve module cover, and pump
platen.
2. Lift the remaining disposable set components from the tub and drape
them over the device with the waste bag on the left side of the device and
the bowl placed loosely in the centrifuge well.
3. Thread the pump tubing around the pump.
4. Install the tubing manifold on the left side of the valve module, pressing it
lightly into place.
5. Floss the tubing into the air detector.
6. Insert the clear tubing and the color-coded lines into the grooves in the
valve module.
7. Close the pump platen.
8. Close and latch the valve module cover.
1.
3.
2.
1. RBC bag
2. Large ratchet clamp
3. Small ratchet clamps