172
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
130
Fluid Detected Early
Explanation:
The bowl optics have detected fluid in bowl earlier than expected for the 225 mL bowl.
NOTE:
This could occur if saline was connected prior to the yellow line being installed in the
valve module and saline inadvertently entered the bowl.
Corrective Action:
If a 225 mL bowl is installed:
1. Check the bowl for proper installation in the centrifuge.
2. Touch
Keep Bowl Size
.
OR
If a 70 mL or 125 mL bowl is installed:
1. Check the bowl and chuck adaptor (70 mL bowl) for proper installation in the centrifuge.
2. Touch
Change Bowl Size
.
3. Select the new bowl size or scan the processing set barcode.
131
Reservoir Load Cell Calibration Failure
Explanation:
•The reservoir load cell calibration values are out of the acceptable range.
•Retry calibration.
Corrective Action:
See service manual for additional information.
132
Waste Bag Load Cell Calibration Failure
Explanation:
•The waste bag load cell calibration values are out of the acceptable range.
•Retry calibration.
Corrective Action:
See service manual for additional information.