Chapter 11, Troubleshooting
169
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
123
Bowl Optics Failure
Explanation:
An issue has been detected with the bowl optics. The power-on self-test has paused. The
procedure may be run, but the user must manually start the Wash phase.
Corrective Action:
To proceed manually:
1. Touch
OK
to continue with POST and run the procedure.
2. Touch
Wash
to start Wash when there is adequate fluid in the bowl.
NOTE:
The bowl is full when the red cell/supernatant interface reaches a point approximately 1/
4 inch over the shoulder of the bowl. The hematocrit of the product may be reduced if the Wash
phase is started before the bowl is full.
OR
Power down the device.
Service is required to fully resolve this issue.
124
Bowl Optics Failure
Explanation:
An issue has been detected with the bowl optics. Processing has paused. Processing can
continue, but the user must manually start the Wash phase.
Corrective Action:
To proceed manually:
1. Touch
Continue
to continue.
2. Touch
Wash
to start Wash when there is adequate fluid in the bowl.
NOTE:
The bowl is full when the red cell/supernatant interface reaches a point approximately 1/
4 inch over the shoulder of the bowl. The hematocrit of the product may be reduced if the Wash
phase is started before the bowl is full.
OR
To end the procedure:
1. Touch
End Procedure
Service is required to fully resolve this issue.
Table 16, Event Messages
ID #
Event Message Text