Chapter 6, General Operation: Sequestration
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Figure 48, Allowing the biohazard waste bag to hang out of the tray
Power-On
Procedure
When ready to initiate a procedure:
1. Ensure the pump platen lever is closed and the valve module cover and
centrifuge header arm are closed and locked.
2. Close the device cover.
3. Press the power switch located on the rear panel of the device.
The device goes through a series of power-on self-tests and advances to
the
Bowl Selection Screen
.
4. Touch
(
Menu
) and select
Sequestration
from the drop-down list to
select the Sequestration protocol. The device displays the Sequestration
Bowl Selection
screen.
Note: During the power-on self-tests (POST), the device checks the interlocks
for the device and manifold covers, the centrifuge arm and the pump platen. To
avoid event messages, it is recommended that these be closed during POST.
If an event message instructing the user to close one of these occurs and
cannot be cleared, close the specified item and restart the device.
Note: When powered on, the device defaults to the startup mode settings group
as determined in the System screen (See
for
more information). To select a different settings group, touch
(
Menu
)
,
select
Settings
from the drop-down menu, and choose the desired settings
group.