Chapter 11, Troubleshooting
193
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
240
Pump Platen Open
Explanation:
The pump platen is not fully closed.
Corrective Action:
1. Open the pump platen.
2. Ensure correct tubing placement around the pump.
3. Close the pump platen.
4. Close the device cover.
If the problem persists, service is required.
241
Fluid Not Detected When Expected
Explanation:
The line sensor did not sense fluid when expected.
Corrective Action:
1. Check the effluent line for correct placement in the effluent line sensor.
2. Clean the effluent line sensor.
3. Check the line for closed clamps, kinks, or occlusions.
4. If using external vacuum, ensure the level is not greater than 250 mmHg.
5. Touch
Continue
to continue.
NOTE:
When using regulated external suction, ensure the A&A line is clamped while setting the
regulator vacuum level to the desired level. If the A&A line is not clamped the vacuum level may
exceed the selected vacuum level and affect performance of the device.
242
Load Disposable
Explanation:
A disposable must be loaded to begin a procedure.
Corrective Action:
1. Load a disposable.
2. Touch
Start Procedure
when ready to begin a procedure.
Table 16, Event Messages
ID #
Event Message Text