ICS-2100 Installation Instructions
84
Doc. 065294-01 3/09
1. Orient the EluGen
cartridge with the fittings
facing up and remove the
plugs from the cartridge
INLET
and
OUTLET
Figure 84. EluGen Cartridge Fitting Plugs
2. Set the EluGen cartridge in the service area on top of the ICS-2100
and orient it with the cable toward the EGC holder (see
Figure 85. EluGen Cartridge in Service Area
OUTLET
INLET
EGC Cable