Setting Up the Eluent Generator
Doc. 065294-01 3/09
87
10. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT
line from port
P
(
2
)
on the injection valve.
11. Use a 10-32 to 10-32 coupler (P/N 042627) to connect the
TO INJ
VALVE IN - P/ELUENT OUT
line to one end of the yellow 0.5 mL/min,
7 MPa (1000 psi) backpressure coil (P/N 053765) (see
12. Direct the free end of the backpressure coil to a temporary waste
container.
Installing the EluGen Cartridge Vent Line
1. Locate the EluGen Cartridge Vent Line Tubing Kit
(P/N 063092) supplied with the EluGen cartridge.
2. Slide the label onto one end of the tubing and then push the barbed
luer lock onto the tubing.
Figure 89. Plumbing Connections for Conditioning the
EluGen Cartridge with Carbonate Eluent
EGC Cable
Temporary
Waste
Container
Back-
pressure
Coil
TO INJ VALVE
IN - P/
ELUENT OUT