Items Required to Complete the Setup Procedure
In addition to the EluGen cartridge and the EGC CO
3
Mixer, you will
need the following items to complete the eluent generator setup. Except
as noted, all items are supplied in the ICS-2100 Ship Kit (P/N 064375).
Item
P/N
Use To
Yellow 0.5 mL/min, 7 MPa
(1000 psi) backpressure coil
053765
Provide backpressure during
setup procedures.
EluGen Cartridge Vent Line
Tubing Kit (supplied with the
cartridge)
063092
Vent the cartridge during
operation.
EGC CO
3
Mixer mounting
clips
063642
Install the mixer.
Phillips head screwdriver (not
supplied)
n/a
Install the EGC CO
3
Mixer
mounting clips.
35 cm (14 in) of black
0.25-mm (0.010-in) ID PEEK
tubing
064346
Connect the EGC degas eluent
outlet and the EGC CO
3
Mixer
inlet.
60 cm (24 in) of black
0.25-mm (0.010-in) ID PEEK
tubing
064343
Connect the EGC CO
3
Mixer
outlet to port
P
(
2
) on the
injection valve.
Temporary waste container
(beaker, bottle, etc.) (not
supplied)
n/a
Collect waste during setup
procedures.
10-32 fittings
10-32 ferrules
043275
043276
Complete the plumbing
connections.
1/8-in fittings
1/8-in ferrules
052267
048949
Complete the plumbing
connections.
10-32 to 10-32 couplers
042627
Complete the plumbing
connections.