ICS-2100 Installation Instructions
88
Doc. 065294-01 3/09
3. Unscrew and remove
the plug from the vent
opening on the side of
the EluGen cartridge
(see
Figure 90. EluGen Cartridge Vent Plug
4. Screw the luer adapter
onto the vent opening.
Remove the cap on
the luer adapter by
twisting.
Save the
cap.
5. Twist the luer lock
and EluGen cartridge
vent line into the luer
adapter (see
Figure 91. Installing the EluGen Cartridge Vent Line
NOTE To prevent leaks from the EluGen cartridge, always
replace the luer adapter cap before moving the
EluGen cartridge.
6. Direct the EluGen cartridge vent line to the rear of the ICS-2100 and
snap it onto one of the tubing clips on the rear panel.
7. Direct the vent line to the waste container and insert it down into the
waste, gas separator tube about 10 to 15 cm (4 to 6 in).
Vent Plug
(remove)
Luer Adapter
EGC
Vent Line
Luer Lock