Setting Up the Eluent Generator
Doc. 065294-01 3/09
61
7. Set the EluGen cartridge in the service area on top of the ICS-2100
and orient it with the cable toward the EGC holder (see
8. Pull the coiled black
EGC IN
and
EGC OUT
lines out from inside the
EGC holder. Disconnect the
EGC IN
line from the coupler that
connects it to the
EGC OUT
line.
9. Connect the
EGC IN
line
to the EluGen
cartridge
INLET
port
10. Remove the coupler
and connect the
EGC
OUT
line to the EluGen
cartridge
OUTLET
port.
Figure 62. EluGen Cartridge Inlet and Outlet Port Connections
11. Turn the EluGen cartridge over (with the fittings facing down).
Shake
the EluGen cartridge vigorously, and tap it with the palm of your
Figure 61. EluGen Cartridge in Service Area
EGC Cable