ICS-2100 Installation Instructions
100
Doc. 065294-01 3/09
21.
Connecting the Analog Output (Optional)
The analog output connector on the IC system rear panel outputs an analog
voltage signal proportional to the conductivity measured by the cell. The selected
data rise time filters the output. The analog output can be connected to an analog-
to-digital converter such as an integrator or other recording device. Refer to the
documentation for the device for connection instructions.
Analog output is 0 to 1V, scaled to the selected range.
For instructions on how to set up the IC system for stand-alone operation with an
AS autosampler, refer to the TTL appendix in the system operator’s manual.
22.
Pressurizing the Eluent Reservoir (Optional)
The IC system does not require pressurized reservoirs. However, if eluent is
manually degassed or is sensitive to contamination, Dionex recommends
pressurizing the reservoir with helium or nitrogen. The air regulator accessory
bracket and other items needed for pressurizing the eluent reservoir must be
ordered separately (P/N 060054).
The following required parts must be ordered separately:
•
Barbed fitting (P/N 030071)
•
Pipe thread reducer (P/N 030087)
•
3-mm (1/8-in) ID tubing (P/N 040793)
•
Air regulator accessory (P/N 060054)