ICS-2100 Installation Instructions
46
Doc. 065294-01 3/09
6.
Locate the yellow
0.5 mL/min, 7 MPa
(1000 psi)
backpressure coil
(P/N 053765) in the
ICS-2100 Ship Kit
(P/N 064375).
7.
Connect one end of
this coil to the EGC
OUTLET
the other end of the
coil unconnected.
Figure 46. EGC Outlet Connection for
Conditioning the EGC
NOTE The backpressure coil connection is temporary; it is
required only during the EluGen cartridge
conditioning procedure described in
the EluGen Cartridge” on page 50
8.
Turn the EluGen cartridge over (with the fittings facing down).
Shake
the EluGen cartridge vigorously, and tap it with the palm of your
hand 10 to 15 times. Check that all bubbles trapped in the eluent
generation chamber are dislodged.
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), a corrosive acid (MSA), or a concentrated
K
2
CO
3
solution. Wear protective eyewear and gloves when handling
the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), un acide corrosif (MSA), ou une solution
concentrée de K
2
CO
3
. Porter des lunettes et des gants protectives en
manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), eine korrodierende Säure (MSA) oder eine konzentrierte
K
2
CO
3
-Lösung. Tragen Sie daher beim Umgang mit Kassette eine
Schutzbrille und Handschuhe.
Backpressure
OUTLET
Port
Coil