ICS-2100 Installation Instructions
36
Doc. 065294-01 3/09
3. Open the priming valve by turning it one-quarter to one-half turn
counterclockwise.
4. On the IC system
panel on the Chromeleon panel tabset, click the
Pump Settings
button.
The
Pump Settings
window appears (see
5. Under
Eluent Flow Valve
, click
Open
.
6. Draw the syringe back to begin pulling eluent through the flow path.
It may take several syringe draws to remove all air or previous eluent
from the tubing. Draw out a minimum of 20 mL of eluent to ensure
that the vacuum degas assembly (if installed) is primed.
7. After priming the lines thoroughly, close the priming valve.
Do not
overtighten the priming valve.
8. Under
Eluent Flow Valve
,
click
Closed
.
Figure 34. Pump Settings Window