Setting Up the Eluent Generator
Doc. 065294-01 3/09
55
Items Required to Complete the Setup Procedure
The following items are required to setup the EluGen cartridge, EPM, and
EGC CO
3
Mixer. Except as noted, all items are supplied in the ICS-2100
Ship Kit (P/N 064375).
Item
P/N
Used To
EPM mounting bracket
063624
Mount the EPM on the EGC holder.
35 cm (14 in) of black 0.25-mm
(0.010-in) ID PEEK tubing
064340
Lengthen the line between the EGC degas
eluent inlet and the EPM
ELUENT OUT
port.
30 cm (12 in) of black 0.25-mm
(0.010-in) ID PEEK tubing
064334
Lengthen the line between the EluGen
cartridge outlet and the EPM
ELUENT IN
port.
28 cm (11 in) of clear 1.58-mm
(0.062-in) ID PTFE tubing
064337
Lengthen the line between the EGC degas
regenerant inlet and the EPM
REGEN OUT
port.
EluGen Cartridge Vent Line
Tubing Kit (supplied with the
cartridge)
063092
Vent the cartridge during operation.
Yellow 0.5 mL/min, 7 MPa
(1000 psi) backpressure coil
053765
Provide backpressure during setup
procedures.
EGC CO
3
Mixer mounting clips
063642
Install the mixer.
Phillips head screwdriver
(not supplied)
n/a
Install the EGC CO
3
Mixer mounting clips.
35 cm (14 in) of black 0.25-mm
(0.010-in) ID PEEK tubing
064346
Connect the EGC degas eluent outlet and the
EGC CO
3
Mixer inlet.
60 cm (24 in) of black 0.25-mm
(0.010-in) ID PEEK tubing
064343
Connect the EGC CO
3
Mixer outlet to port
P
(
2
) on the injection valve.
Temporary waste container
(beaker, bottle, etc.) (not supplied)
n/a
Collect waste during setup procedures.
10-32 to 10-32 coupler
10-32 to 1/4-28 coupler
042627
042806
Complete the connections.
10-32 fittings
10-32 ferrules
043275
043276
Complete the connections.
1/8-in fittings
1/8-in ferrules
052267
048949
Complete the connections.