ICS-2100 Installation Instructions
102
Doc. 065294-01 3/09
22.2
Applying the Pressure
1. Before pressurizing, make sure the reservoir is filled and connected to
the system (see
2. Pull out the air regulator knob and turn it fully counterclockwise to
ensure there is no pressure to the reservoir when the gas source is
turned on.
3. Turn on the gas source.
4. Pull out the air regulator knob and turn it clockwise to pressurize the
eluent reservoir.
5. Adjust the pressure to between 0.03 and 0.04 MPa (5 and 6 psi). Push
the regulator knob back in.
Do not pressurize the eluent reservoir above 0.07 MPa (10 psi).
Ne mettez jamais les réservoirs d'éluants sous une pression
supérieure à 0,07 MPa (10 lb/po²).
Setzen Sie den Eluentbehälter auf keinen Fall einem Druck über
0,07 MPa aus.