Setting Up the Eluent Generator
Doc. 065294-01 3/09
51
15. Disconnect the
backpressure coil
from the EluGen
cartridge
OUTLET
port
.
16. Connect the
EGC
OUT
line to the
OUTLET
Figure 52. EGC Inlet and Outlet Connections
17. Turn over the EluGen cartridge. Check for bubbles; if necessary,
shake and tap the EluGen cartridge to remove bubbles.
18. Reinstall the EluGen cartridge in the holder.
19. Do not reconnect the EluGen cartridge cable until after hydrating the
CR-TC.
Hydrating the CR-TC
Always hydrate the CR-TC before initial operation. This requires running
ASTM filtered, Type 1 (18 megohm-cm) deionized water through the CR-
TC for 10 minutes, while bypassing the columns and suppressor.