ICS-2100 Installation Instructions
74
Doc. 065294-01 3/09
3. Connect the pump outlet to the mixer inlet (see
4. Direct the mixer outlet line to a temporary waste container, such as a
beaker.
5. Verify that the suppressor is off.
6. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2100 touch screen
SUPPRESSOR
page. Disabling
this option lets you turn on the pump, without automatically turning
on the suppressor.
7. Set the pump flow rate to
5.00 mL/min.
8. Turn on the pump.
9. Run the pump until the mixer is filled and there is a consistent flow of
water exiting the mixer outlet.
It takes about 5 minutes to fill a 4-mm mixer and about 2 minutes to
fill a 2-mm mixer.
10. Turn off the pump.
Figure 77. Pump Outlet Connected to Mixer Inlet for Mixer Hydration
To avoid damaging the suppressor, always turn off the suppressor
before filling the mixer. The pump flow is on during filling, however,
no flow reaches the suppressor.
Mixer Inlet
Pump Outlet