Connecting an Autosampler (Optional)
Doc. 065294-01 3/09
7
5. If the coupler is not visible, grasp the green waste line where it exits the
tubing chase and pull it toward the front of the system until the coupler is
outside the chase.
6. Disconnect the green waste line (labeled
TO AS50 TRAY
) from the coupler and
route the line to the drip tray in the autosampler compartment.
You can either leave the disconnected PTFE waste line and coupler in the
tubing chase, or pull the line out from the front.
7. Bend the green waste line about 5 mm (0.2 in) from the end (to ensure it stays
in the AS drip tray). When bending the tubing, be sure not to restrict the inner
diameter.
8. Push the end of the green waste line into the small round opening on the right
side of the AS tray (see
NOTE After the AS is configured for use with the ICS-2100
), all AS injection types are
supported. In addition, all injection valve
commands (whether issued from Chromeleon or the
AS front panel) can control the injection valve.
Make sure the injection valve waste line is not elevated above the
injection valve at any point between the valve and the autosampler
drip tray.