Setting Up the Eluent Reservoir
Doc. 065294-01 3/09
33
15.
Setting Up the Eluent Reservoir
1. Rinse the eluent reservoir with ASTM filtered, Type I (18-megohm)
deionized water.
2. Fill the reservoir with ASTM filtered, Type I (18-megohm) deionized water.
3. Place the reservoir on top of the IC system.
4. Install the cap and handtighten.
5. Connect the liquid lines from the IC system to the corresponding lines exiting
the eluent bottle.
NOTE The IC system does not require pressurized reservoirs.
However, if eluent is manually degassed or is sensitive to
contamination, Dionex recommends pressurizing the
reservoir with helium or nitrogen. See
for
instructions.
VERSCHLIESSEN SIE DEN ABFALLBEHÄLTER NICHT: Der ICS-2100 ,
Atlas Electrolytic Suppressor und Self-Regenerating Suppressor
(SRS) verwenden einen Elektrolyseprozess, wodurch kleine Mengen
an Sauerstoff und Wasserstoff entstehen. Führen Sie einen
schwarzen Gasabscheiderschlauch (ID = 1,3 cm; Bestell-Nr. 045460) in
einen offenen (unverschlossenen) Abfallbehälter, damit sich das Gas
nicht in einem geschlossenen Behälter sammelt und aufkonzentriert.
Verbinden Sie die mit Waste, Gas Separator bezeichnete Leitung mit
dem Abfallschlauch.