Pressurizing the Eluent Reservoir (Optional)
Doc. 065294-01 3/09
101
22.1
Connecting the Gas Source
illustrates the eluent reservoir after all gas connections are
complete.
1. Use the barbed fitting and pipe thread reducer to connect the 3-mm
(1/8-in) ID tubing to a clean gas source regulated to between 0.55 and
0.83 MPa (80 and 120 psi).
2. Push the other end of the tubing onto the inlet of the air regulator.
3. Remove the eluent reservoir cap.
4. Connect the gas line from the eluent reservoir cap to the gas line on
the air regulator outlet.
5. Slip the air regulator accessory bracket over the neck of the eluent
Figure 101. IC System Pressurized Eluent Reservoir Connections (Optional)
Gas Inlet Line
3-mm (1/8-in) ID
Eluent Gas
Line
Air Regulator
Accessory and
Bracket