ICS-2100 Installation Instructions
6
Doc. 065294-01 3/09
4. Connect the needle seal line to port
S (5)
) connects the green waste line from port
W (6)
to a
3-mm (1/8-in) ID PTFE waste line. The waste line is routed through the
tubing chase to the IC system rear panel.
The needle seal line from the AS autosampler inject port to the IC
system injection valve is specially constructed. It consists of two
tubes (one inside the other) that are sealed at the end. This seal must
be maintained.
In addition, the line is calibrated for use with the IC system. To ensure
correct autosampler function, this line MUST NOT BE CUT! If
replacement is required, refer to the AS operator’s manual for the
required part number and replacement instructions.
Figure 7. Injection Valve
Figure 8. Injection Valve Waste Line Coupler
Autosampler
Needle Seal
Line
Waste
Line
Coupler
Green
Waste
Line