ICS-2100 Installation Instructions
52
Doc. 065294-01 3/09
illustrates the liquid flow path through the CR-TC during the
hydration procedure.
1. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT
line from port
P
(
2
)
on the injection valve.
2. Disconnect the
TO SRS/AES REGEN OUT
line from the suppressor.
3. Use a 10-32 to 1/4-28 coupler (P/N 042806) to connect the
TO INJ
VALVE IN - P/ELUENT OUT
line to the
SRS/AES REGEN OUT
line.
4. Verify that the suppressor is off.
5. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2100 touch screen
SUPPRESSOR
page. Disabling
Figure 53. CR-TC Plumbing for Hydration
To avoid damaging the suppressor, always turn off the suppressor
before hydrating the CR-TC. The pump flow is on during conditioning,
however, no flow reaches the suppressor.