Setting Up the Eluent Generator
Doc. 065294-01 3/09
41
illustrates the liquid flow path through the eluent generator
components after all setup steps are complete and the eluent generator is
ready for operation.
Figure 37. Eluent Generator Flow Schematic with CR-TC:
Use for Generating KOH, LiOH, MSA, or NaOH Eluent
El
ue
n
t In
Re
ge
n
In
Re
ge
n
Out
Eluent
Out
CR-TC
EGC
DEGAS
EGC
IN
Connect to
Gas Separator
Assembly
TO DEGAS
ELUENT OUT
TO CR-TC
REGEN OUT
(blue)
TO SRS/AES
REGEN OUT
TO
C
R-T
C
RE
GE
N I
N
(o
ran
ge
)
WASTE,
GAS
SEPARATOR
DEGAS
ELUENT OUT
TO INJ
VALVE IN - P/
ELUENT OUT
Connect to
Inject Valve
Port P (2)
Connect to
Suppressor
Regen Out
Pre-plumbed
Pump Pulse
Damper
to
EGC II
KOH, LiOH,
MSA, or NaOH
Inlet
Outlet
EGC VENT
Send to waste
TO DEGAS
ELUENT IN
EGC
OUT
TO CR-TC
ELUENT IN
(red)
TO CR-TC/EPM
ELUENT OUT
(yellow)
TO PUMP/
DAMPER
*Add backpressure coils, if
needed, to create a system
backpressure of between 14 and
16 MPa (2000 and 2300 psi).
Back-
pressure
Coil*