ICS-2100 Installation Instructions
82
Doc. 065294-01 3/09
illustrates the liquid flow path through the eluent generator
components after all setup steps are complete and the eluent generator is
ready for operation.
Figure 82. Eluent Generator Flow Schematic with an EGC CO
3
Mixer:
Use for Generating Carbonate Eluent
EGC
DEGAS
EGC
IN
Connect to
Gas Separator
Assembly
TO SRS/AES
REGEN OUT
WASTE,
GAS
SEPARATOR
Pre-plumbed
Pump Pulse
Damper
to
Inlet
Outlet
EGC VENT
Send to waste
EGC
OUT
EGC II
K CO
2
3
TO DEGAS
ELUENT IN
Back-
pressure
Coil*
Connect
Inject Valve
Port P (2)
to
EGC CO
Mixer
3
Connect to
Suppressor
Regen Out
TO DEGAS
ELUENT OUT
*Add backpressure coils, if needed, to
create a system backpressure of between
14 and 16 MPa (2000 and 2300 psi).
TO PUMP/
DAMPER
TO EGC CO3
MIXER IN
TO EGC CO3
MIXER OUT
TO INJ VALVE
PORT P
TO INJ
VALVE IN - P
ELUENT OUT