ICS-2100 Installation Instructions
106
Doc. 065294-01 3/09
sample processing parameters (the sample name, sample type, injection
volume, etc.).
•
After creating the sequence, you can start sample processing. On the
Sequence Control
panel, click
Load Sequence
and then click
Start Batch
.
25.
Troubleshooting the USB Configuration Error
Problem:
When the IC system power is turned on for the first time, a Windows
message box appears asking for a USB configuration file (cmwdmusb.inf).
Possible Cause:
The USB cable was connected and the power turned on before
Chromeleon was installed.
Solution:
Follow these steps to resolve the problem:
1. Click
Cancel
in the Windows message box.
2. Turn off the power to the USB device and unplug the USB cable from the
computer.
4. Reconnect the USB cable to the computer and turn on the power to the
device. Windows will now automatically recognize the USB device.