Setting Up the Eluent Generator
Doc. 065294-01 3/09
93
Filling the EGC CO
3
Mixer with the Carbonate Eluent
Before initial use, follow the steps in this section to fill the EGC CO
3
Mixer with the carbonate eluent required for the application to be run.
illustrates the liquid flow path through the eluent generator
components during initial filling of the EGC CO
3
Mixer with carbonate
eluent.
Figure 96. Plumbing for Filling the EGC CO
3
Mixer with Carbonate Eluent
EGC
DEGAS
EGC
IN
Connect to
Gas Separator
Assembly
TO DEGAS
ELUENT OUT
TO SRS/AES
REGEN OUT
WASTE,
GAS
SEPARATOR
Connect to
Suppressor
Regen Out
Pre-plumbed
Pump Pulse
Damper
to
Inlet
Outlet
EGC VENT
Send to waste
EGC
OUT
EGC II
K CO
2
3
TO DEGAS
ELUENT IN
Back-
pressure
Coil
TO INJ
VALVE IN - P/
ELUENT OUT
To waste
EGC CO
Mixer
3
TO EGC CO3
MIXER IN
TO EGC CO3
MIXER OUT
14 MPa (2000 psi)
at 1.0 mL/min
TO PUMP/
DAMPER