ICS-2100 Installation Instructions
34
Doc. 065294-01 3/09
16.
Entering the Eluent Level
After filling the reservoir, enter the volume of liquid in the reservoir.
1. On the IC system
panel on the Chromeleon panel tabset, click the
Pump
Settings
button.
The
Pump Settings
window appears (see
2. Use the slider to enter the volume (in liters), or enter the volume directly into
the
Eluent Bottle
box.
The IC system determines the eluent usage by monitoring the flow rate and the
length of time the pump is on, and updates the
Eluent Bottle
volume as the eluent
is used up. A warning appears if the level falls below 200 mL. Warnings are
repeated at 100 mL and 0 mL.
If preferred
,
you can enter the eluent bottle level on the touch screen
HOME
page,
instead of on the Chromeleon Control panel.
Figure 32. Pump Settings Window: Setting the Eluent Level
In order for the level displayed in the Eluent Bottle box and gauge to
be accurate, you must enter the level when the reservoir is filled. The
IC system does not automatically detect when the reservoir is filled,
nor when it is empty.
Enter the
volume of
liquid in the
reservoir
(in liters)
Note: The Eluent
Bottle field and
slider gauge are
updated as the
liquid is used up.