Setting Up the Eluent Generator
Doc. 065294-01 3/09
59
2. Use the precut tubing assemblies included in the ICS-2100 Ship Kit
to extend the length of three lines as described below:
a.
Disconnect the
TO
CR-TC ELUENT IN
line (red label) from the
coupler that connects it to the
TO CR-TC ELUENT OUT
line (yellow
label). Connect a 32 to 10-32 coupler to the end of the
TO
CR-TC
ELUENT IN
line and connect the 30 cm (12 in) length of black
PEEK tubing (red labels) (P/N 064334) to the coupler (see
b. Connect the 35 cm (14 in) length of black PEEK tubing (yellow
labels) (P/N 064340) to the coupler on the
TO CR-TC ELUENT OUT
line (yellow label).
c.
Disconnect the
TO CR-TC REGEN IN
line (orange label) from the
coupler that connects it to the
TO CR-TC REGEN OUT
line (blue
label). Connect
the
28 cm (11 in) length of clear PTFE tubing
(blue labels) (P/N 064337) to the coupler on the
TO CR-TC REGEN
OUT
line (blue label).
Figure 58. EPM Eluent and Regen Line Extensions
TO CR-TC/EPM
ELUENT OUT
Line (yellow
label)
TO CR-TC/EPM
ELUENT IN
Line (red label)
TO CR-TC/EPM
REGEN OUT
Line (blue label)